Ventec Life Systems, Inc. — VOCSN Model Number: PRT-01100-000, Package PN:PRT-01126-000. used for ventilatory support.

Official product recall notices and clearly labeled enforcement reference records.

Reference record — not a live warning · enforcement-reference · official-source

Ventilator's Internal Flow transducer cable connector may disconnect from internal board, potentially resulting in system reboot and interruption of ventilation therapy.

Posted by source
Source last updated
Current version retrieved
Provider
FDA — device enforcement reference
Location
Bothell, WA, US
hazard
Ventilator's Internal Flow transducer cable connector may disconnect from internal board, potentially resulting in system reboot and interruption of ventilation therapy.
notice
Periodic enforcement reference; not a real-time safety alert or a reliable recall-lifecycle tracker.
product
VOCSN Model Number: PRT-01100-000, Package PN:PRT-01126-000. used for ventilatory support.
distribution
Distribution US nationwide including Puerto Rico.
event number
99835
manufacturer
Ventec Life Systems, Inc.
recall number
Z-0126-2027
source status
Open, Classified
freshness minutes
180
Read the official source

Source use policy · openFDA reference data; not a real-time public alert service

Update history (1 recent revisions)

Report a correction · Related news · Fire information · Clips