Steris Corporation — VERIFY Biological/Integrator Challenge Pack Model/Catalog Number: LCB006
Official product recall notices and clearly labeled enforcement reference records.
Reference record — not a live warning · enforcement-reference · official-source
Geobacillus stearothermophilus spore population in affected products may fall below minimum shelf life specification. When used to monitor steam sterilization cycles, the potential hazard is that the biological indicator will show no growth, indicating that the steam cycle was effective, even if the steam cycle did not achieve the intended spore reduction
- Posted by source
- Source last updated
- Current version retrieved
- Provider
- FDA — device enforcement reference
- Location
- Mentor, OH, US
- hazard
- Geobacillus stearothermophilus spore population in affected products may fall below minimum shelf life specification. When used to monitor steam sterilization cycles, the potential hazard is that the biological indicator will show no growth, indicating that the steam cycle was effective, even if the steam cycle did not achieve the intended spore reduction
- notice
- Periodic enforcement reference; not a real-time safety alert or a reliable recall-lifecycle tracker.
- product
- VERIFY Biological/Integrator Challenge Pack Model/Catalog Number: LCB006
- distribution
- Worldwide - US Nationwide distribution.
- event number
- 99704
- manufacturer
- Steris Corporation
- recall number
- Z-0070-2027
- source status
- Open, Classified
- freshness minutes
- 180