Philips North America — Ingenia Elition S and Ingenia Elition X.; Model number (REF): 1. 781357 (S), 2. 782106 (S), 3. 782137 (S), 4. 782157 (S), 5. 781358 (X), 6. 782107 (X), 7. 782119 (X), 8. 782136 (X)

Official product recall notices and clearly labeled enforcement reference records.

Reference record — not a live warning · enforcement-reference · official-source

When exporting diffusion tensor imaging (DTI) scans from a MR system using the Enhanced MR DICOM format, the exports contain incorrect data (diffusion directions). If the Enhanced MR DICOM export is used in conjunction with third-party software for tractography (e.g. for mapping nerve fiber pathways in the brain), the software will not be able to correctly process the DTI data for reconstruction.

Provider
FDA — device enforcement reference
Source observed
8/28/2026, 12:00:00 AM UTC
Source updated
9/25/2026, 12:00:00 AM UTC
Location
Cambridge, MA, US
hazard
When exporting diffusion tensor imaging (DTI) scans from a MR system using the Enhanced MR DICOM format, the exports contain incorrect data (diffusion directions). If the Enhanced MR DICOM export is used in conjunction with third-party software for tractography (e.g. for mapping nerve fiber pathways in the brain), the software will not be able to correctly process the DTI data for reconstruction.
notice
Periodic enforcement reference; not a real-time safety alert or a reliable recall-lifecycle tracker.
product
Ingenia Elition S and Ingenia Elition X.; Model number (REF): 1. 781357 (S), 2. 782106 (S), 3. 782137 (S), 4. 782157 (S), 5. 781358 (X), 6. 782107 (X), 7. 782119 (X), 8. 782136 (X), 9. 782158 (X), 10. 782198 (X); Software Versions: R11, R12 and R13;
distribution
Worldwide distribution - Nationwide and the countries of Algeria, Argentina, Australia, Austria, Bangladesh, Belgium, Brazil, Canada, Chile, China, Colombia, Cyprus, Czech Republic, Denmark, Dominican Rep, Ecuador, Egypt, Equator. Guinea, Finland, France, French Guiana, Georgia, Germany, Ghana, Greece, Honduras, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Kyrgyzstan, Latvia, Lebanon, Luxembourg, Malaysia, Malta, Martinique, Mauritius, Mexico, Moldova, Monaco, Morocco, Myanmar, Namibia, Nepal, Netherlands, New Zealand, Nicaragua, Norway, Pakistan, Palestine, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russian Fed., Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, South Korea, Spain, Sweden, Switzerland, Syria, Taiwan, Thailand, Tunisia, T¿rkiye, United Kingdom, Uruguay, Utd.Arab.Emir., Uzbekistan, Vietnam.
event number
99834
manufacturer
Philips North America
recall number
Z-3311-2026
source status
Open, Classified
freshness minutes
180
Read the official source

Source use policy · openFDA reference data; not a real-time public alert service

Update history (1 recent revisions)
  • 10/8/2026, 2:10:36 AM UTC