Percussionaire Corporation — Travel Air, Product Number: S00055; Phasitron 5, Product Number: P5 used with Instruction Manual
Official product recall notices and clearly labeled enforcement reference records.
Reference record — not a live warning · enforcement-reference · official-source
Labeling updated warning not to use Intrapulmonary percussive ventilation (IPV) in conjunction with volatile anesthetic delivery systems because IPV may alter controlled delivery of volatile anesthetic agents, resulting in deeper/lighter anesthetic effects that may result in hypertonia/hypotonia, tachycardia/ bradycardia, elevated intracranial pressure, awareness during anesthesia, oversedation
- Provider
- FDA — device enforcement reference
- Source observed
- 8/18/2026, 12:00:00 AM UTC
- Source updated
- 9/28/2026, 12:00:00 AM UTC
- Location
- Sandpoint, ID, US
- hazard
- Labeling updated warning not to use Intrapulmonary percussive ventilation (IPV) in conjunction with volatile anesthetic delivery systems because IPV may alter controlled delivery of volatile anesthetic agents, resulting in deeper/lighter anesthetic effects that may result in hypertonia/hypotonia, tachycardia/ bradycardia, elevated intracranial pressure, awareness during anesthesia, oversedation
- notice
- Periodic enforcement reference; not a real-time safety alert or a reliable recall-lifecycle tracker.
- product
- Travel Air, Product Number: S00055; Phasitron 5, Product Number: P5 used with Instruction Manual
- distribution
- Worldwide - US Nationwide distribution including in the states of CA, PA, OK, GA, TX, OH, WI, WA, NY, MO, FL, DE, LA, SC, WV, NJ, NE, KS, NC, AL, CO, VA, OR, AZ, IL, CT, ID, NH, UT, TN, MD, VT, MS, NV, DC, ME, IA, AR, HI, ND, IN, MT, NM, MN, MI, SD, MA, AK, RI and the countries of Canada, Belgium, Chile, France, Germany, Israel, Italy, Latvia, Netherlands, Norway, Oman, Philippines, Portugal, Qatar, Russia, Saudi Arabia, Spain, Sweden, Switzerland, Turkey, United Arab Emirates, United Kingdom, Kuwait, Australia, South Africa.
- event number
- 99617
- manufacturer
- Percussionaire Corporation
- recall number
- Z-3324-2026
- source status
- Open, Classified
- freshness minutes
- 180
Source use policy · openFDA reference data; not a real-time public alert service
Update history (1 recent revisions)
- 10/8/2026, 2:10:43 AM UTC
Report a correction · Related news · Fire information · Clips