Medtronic MiniMed, Inc. — CareLink Clinic, REF: MMT-7350

Official product recall notices and clearly labeled enforcement reference records.

Reference record — not a live warning · enforcement-reference · official-source

CareLink Clinic software version 4.4A includes a software defect in the Settings Calculator feature in which the calculation logic fails to use custom values entered by healthcare professionals (HCPs) for Total Daily Dose (TDD) or Long-acting insulin when calculating the Carb Ratio, Insulin Sensitivity Factor (ISF), and Max Recommended value for Rapid-acting insulin. Instead, the calculator continues to use the system-recommended values, resulting in incorrect rapid-acting insulin parameters. If an HCP or patient inputs these incorrect setting values into the patient's MiniMed Go app, the patient could deliver higher or lower insulin doses compared to their insulin needs, potentially leading to over/under delivery of insulin with a severity ranging from Catastrophic (death from severe hypoglycemia) to Minor (self-treatable hypoglycemia/hyperglycemia).

Posted by source
Source last updated
Current version retrieved
Provider
FDA — device enforcement reference
Location
Northridge, CA, US
hazard
CareLink Clinic software version 4.4A includes a software defect in the Settings Calculator feature in which the calculation logic fails to use custom values entered by healthcare professionals (HCPs) for Total Daily Dose (TDD) or Long-acting insulin when calculating the Carb Ratio, Insulin Sensitivity Factor (ISF), and Max Recommended value for Rapid-acting insulin. Instead, the calculator continues to use the system-recommended values, resulting in incorrect rapid-acting insulin parameters. If an HCP or patient inputs these incorrect setting values into the patient's MiniMed Go app, the patient could deliver higher or lower insulin doses compared to their insulin needs, potentially leading to over/under delivery of insulin with a severity ranging from Catastrophic (death from severe hypoglycemia) to Minor (self-treatable hypoglycemia/hyperglycemia).
notice
Periodic enforcement reference; not a real-time safety alert or a reliable recall-lifecycle tracker.
product
CareLink Clinic, REF: MMT-7350
distribution
US Nationwide distribution in the state of KS.
event number
99765
manufacturer
Medtronic MiniMed, Inc.
recall number
Z-0203-2027
source status
Open, Classified
freshness minutes
180
Read the official source

Source use policy · openFDA reference data; not a real-time public alert service