Howmedica Osteonics Corp. — TS Misc. Instruments Tray, Triathlon. Catalog Number: 6543-6-519.

Official product recall notices and clearly labeled enforcement reference records.

Reference record — not a live warning · enforcement-reference · official-source

Reprocessing instructions provided in LSTPI-B specify a sterilization dry time of 30 minutes for the trays sterilized in blue wrap. When blue wrap is used, the 30-minute dry time may lead to a small amount of residual moisture remaining after sterilization.

Posted by source
Source last updated
Current version retrieved
Provider
FDA — device enforcement reference
Location
Mahwah, NJ, US
hazard
Reprocessing instructions provided in LSTPI-B specify a sterilization dry time of 30 minutes for the trays sterilized in blue wrap. When blue wrap is used, the 30-minute dry time may lead to a small amount of residual moisture remaining after sterilization.
notice
Periodic enforcement reference; not a real-time safety alert or a reliable recall-lifecycle tracker.
product
TS Misc. Instruments Tray, Triathlon. Catalog Number: 6543-6-519.
distribution
These trays are provided by sale or loan. Loans are provided through Distributor. Foreign Consignees will be identified by the countries impacted. 158 trays have been distributed OUS.
event number
99883
manufacturer
Howmedica Osteonics Corp.
recall number
Z-0167-2027
source status
Open, Classified
freshness minutes
180
Read the official source

Source use policy · openFDA reference data; not a real-time public alert service

Update history (1 recent revisions)

Report a correction · Related news · Fire information · Clips