Reprocessing instructions provided in LSTPI-B specify a sterilization dry time of 30 minutes for the trays sterilized in blue wrap. When blue wrap is used, the 30-minute dry time may lead to a small amount of residual moisture remaining after sterilization.
- Posted by source
- Source last updated
- Current version retrieved
- Provider
- FDA — device enforcement reference
- Location
- Mahwah, NJ, US
- hazard
- Reprocessing instructions provided in LSTPI-B specify a sterilization dry time of 30 minutes for the trays sterilized in blue wrap. When blue wrap is used, the 30-minute dry time may lead to a small amount of residual moisture remaining after sterilization.
- notice
- Periodic enforcement reference; not a real-time safety alert or a reliable recall-lifecycle tracker.
- product
- TS Misc. Instruments Tray, Triathlon. Catalog Number: 6543-6-519.
- distribution
- These trays are provided by sale or loan. Loans are provided through Distributor. Foreign Consignees will be identified by the countries impacted. 158 trays have been distributed OUS.
- event number
- 99883