CooperSurgical, Inc. — INCA (Infant Nasal Cannulae Assy) Neonatal Nasal CPAP, Replacement Set. Stay-Flex tubing component. 5 units per box. Model Numbers: 44-0712, 44-0707, 44-0709, 44-0715.

Official product recall notices and clearly labeled enforcement reference records.

Reference record — not a live warning · enforcement-reference · official-source

Potential to exhibit a loose connection between the flexible tubing and the nasal cannula, during assembly or while in use, which may result in loss of CPAP pressure, possibly reducing the effectiveness of treatment and lead to decreased oxygenation until connection is restored.

Provider
FDA — device enforcement reference
Source observed
8/11/2026, 12:00:00 AM UTC
Source updated
9/18/2026, 12:00:00 AM UTC
Location
Trumbull, CT, US
hazard
Potential to exhibit a loose connection between the flexible tubing and the nasal cannula, during assembly or while in use, which may result in loss of CPAP pressure, possibly reducing the effectiveness of treatment and lead to decreased oxygenation until connection is restored.
notice
Periodic enforcement reference; not a real-time safety alert or a reliable recall-lifecycle tracker.
product
INCA (Infant Nasal Cannulae Assy) Neonatal Nasal CPAP, Replacement Set. Stay-Flex tubing component. 5 units per box. Model Numbers: 44-0712, 44-0707, 44-0709, 44-0715.
distribution
Nationwide distribution to CT, FL, GA, IA, IL, IN, MD, MI, MS, NJ, NY, OH, PA, VA. International distribution to Brazil and Peru.
event number
99643
manufacturer
CooperSurgical, Inc.
recall number
Z-3140-2026
source status
Open, Classified
freshness minutes
180
Read the official source

Source use policy · openFDA reference data; not a real-time public alert service

Update history (1 recent revisions)
  • 10/8/2026, 2:10:44 AM UTC

Report a correction · Related news · Fire information · Clips