Beckman Coulter, Inc. — ROTOR ASSY-SW40, Reference Number 331302, when the rotors are installed in the following centrifuges: Optima XPN-100 (Model Numbers A99846 and A94469), Optima XPN-90 (Model Numbers

Official product recall notices and clearly labeled enforcement reference records.

Reference record — not a live warning · enforcement-reference · official-source

Optima XE XPN ultracentrifuge rotors may contain a microstructure defect that can cause rotors to break into two or more pieces (bifurcate) during a centrifugation run, resulting in destruction of samples being processed. When used for in vitro diagnostic (IVD) sample preparation, rotor breakage may delay patient test results. Additionally, sample destruction may expose laboratory personnel to biohazardous materials.

Posted by source
Source last updated
Current version retrieved
Provider
FDA — device enforcement reference
Location
Indianapolis, IN, US
hazard
Optima XE XPN ultracentrifuge rotors may contain a microstructure defect that can cause rotors to break into two or more pieces (bifurcate) during a centrifugation run, resulting in destruction of samples being processed. When used for in vitro diagnostic (IVD) sample preparation, rotor breakage may delay patient test results. Additionally, sample destruction may expose laboratory personnel to biohazardous materials.
notice
Periodic enforcement reference; not a real-time safety alert or a reliable recall-lifecycle tracker.
product
ROTOR ASSY-SW40, Reference Number 331302, when the rotors are installed in the following centrifuges: Optima XPN-100 (Model Numbers A99846 and A94469), Optima XPN-90 (Model Numbers A99842 and A94468), Optima XPN-80 (Model Numbers A99839 and A95765), Optima XE-100 (Model Numbers A99836 and A94516), and Optima XE-90 (Model Numbers A99833 and A94471)
distribution
US Nationwide distribution.
event number
99882
manufacturer
Beckman Coulter, Inc.
recall number
Z-0201-2027
source status
Open, Classified
freshness minutes
180