Beckman Coulter, Inc. — ROTOR ASSY-SW40, Reference Number 331302, when the rotors are installed in the following centrifuges: Optima XPN-100 (Model Numbers A99846 and A94469), Optima XPN-90 (Model Numbers
Official product recall notices and clearly labeled enforcement reference records.
Reference record — not a live warning · enforcement-reference · official-source
Optima XE XPN ultracentrifuge rotors may contain a microstructure defect that can cause rotors to break into two or more pieces (bifurcate) during a centrifugation run, resulting in destruction of samples being processed. When used for in vitro diagnostic (IVD) sample preparation, rotor breakage may delay patient test results. Additionally, sample destruction may expose laboratory personnel to biohazardous materials.
- Posted by source
- Source last updated
- Current version retrieved
- Provider
- FDA — device enforcement reference
- Location
- Indianapolis, IN, US
- hazard
- Optima XE XPN ultracentrifuge rotors may contain a microstructure defect that can cause rotors to break into two or more pieces (bifurcate) during a centrifugation run, resulting in destruction of samples being processed. When used for in vitro diagnostic (IVD) sample preparation, rotor breakage may delay patient test results. Additionally, sample destruction may expose laboratory personnel to biohazardous materials.
- notice
- Periodic enforcement reference; not a real-time safety alert or a reliable recall-lifecycle tracker.
- product
- ROTOR ASSY-SW40, Reference Number 331302, when the rotors are installed in the following centrifuges: Optima XPN-100 (Model Numbers A99846 and A94469), Optima XPN-90 (Model Numbers A99842 and A94468), Optima XPN-80 (Model Numbers A99839 and A95765), Optima XE-100 (Model Numbers A99836 and A94516), and Optima XE-90 (Model Numbers A99833 and A94471)
- distribution
- US Nationwide distribution.
- event number
- 99882
- manufacturer
- Beckman Coulter, Inc.
- recall number
- Z-0201-2027
- source status
- Open, Classified
- freshness minutes
- 180
Source use policy · openFDA reference data; not a real-time public alert service
Update history (1 recent revisions)
Report a correction · Related news · Fire information · Clips